Date: 15 October 2009
Description: Pre-deployment Calibration
Location: Operations Center
Samplers: Brandon Sackmann (ECY)
Mya Keyzers (ECY)
SUNA SN: SATSLF0044
Ref. Spectrum: SUNA044_2009_08_19_FACTORY_CAL.txt (Pre)
SUNA044_2009_08_19_FACTORY_CAL.txt (Post)
Lamp Time: 41.6 hr (96% remaining)
Units: mg/L
Laboratory: Manchester Environmental Laboratory (MEL)
Notes: Ultra-clean de-ionized water (DIW); SUNA-Nitrate values between +/-0.028 mg/L indicate the instrument response has not drifted. MEL results are to be used as a filter blank to ensure that samples are not being contaminated by the filtering apparatus.

SUNA Avg. Nitrate and %RSD are calculated from 50 discrete scans (i.e., n = 50).

U1-3 correspond to Nitrate, Nitrite, and Nitrite+Nitrate, respectively; indicates that the analyte was not detected at or above the reported result.